Gas sensor showing incorrect readings? 5 common causes of measurement errors

A gas sensor is installed and the calibration is correct, yet the readings still fluctuate. What initially looks like a technical fault often has another cause: the environmental conditions.

Temperature, humidity, pressure and the properties of the gas sample can have a considerable effect on measurement accuracy. In industrial applications, where precise gas measurements affect process quality, efficiency and safety, it is worth taking a closer look at possible sources of interference.

1. Temperature: When heat affects the measurement

Gas sensors do not operate independently of their surroundings. Rising or falling temperatures can affect both electronic components and the physical properties of the gas being measured.

In NDIR gas sensors, this can affect the infrared source and the detector, for example. Rapid temperature changes in particular can cause temporary measurement deviations.

The solution: Appropriate temperature compensation reduces these effects. However, the sensor must be designed for the actual temperature range of the application.

Schematic illustration of a blue gas sensor module between heat and cold

2. Humidity: An underestimated source of interference

Water vapour can affect gas measurement in different ways. In optical measurement methods, humidity can interfere in certain absorption bands.

The problem becomes more serious when water vapour condenses. Liquid droplets can affect optical components and block gas pathways.

The solution: Depending on the application, controlled gas conditioning, suitable filters or drying the gas sample can help. These measures must not alter the concentration of the target gas beyond acceptable limits.